HIGH PERFORMANCE INCUBATORS MODELS: 311D, 611D INSTALLATION AND INSTRUCTION MANUAL
Section 7 OPERATION 7.1 Check power supply against unit serial plate. They must match. 7.2 Plug service cord into the electrical outlet. Push pow
11stable temperature the fan may be turned off.
CLEANING AND ROUTINE CARE Section 8 Both the interior and exterior of the incubator should be wiped down with a soft cloth periodically to prevent a
SERVICE AND TROUBLE-SHOOTING Section 9 TEMPERATURE Temp-1 Temperature too high-display and reference thermometer don’t match 1/ Controller
14 1/ Safety set too low – see section 7.5. 2/ Controller set too low – see section 7.3. 3/ Unit not recovered from door opening – wait for di
15including motors, arcing relays or radio transmitters. 7/ Bad connection on temperature sensor or faulty sensor – check connectors for continuit
16 1/ If shaft spins freely: check connections to motor and check voltage to motor. 2/ If shaft rubs or is frozen, relieve binding and retest.
17 OTHER Other-1 Controller on at all times - "locked-up" 1/ Turn unit off and on to reset. 2/ If
18 PARTS LIST Description 120V 220V Blower Motor I-104100 I-210001 Circuit Breaker I-1100505 I-1100505 Convenience Outlet I-100020 I
2TABLE OF CONTENTS SECTION 1.0 INTRODUCTION SECTION 2.0 WARRANTY SECTION 3.0 SPECIFICATIONS SECTION 4.0 RECEIVING
Section 1 INTRODUCTION The Labnet Incubators have been designed for general purpose incubations (<65ºC), as well as those applications requiring h
Section 2 WARRANTY Labnet International, Inc. guarantees that the Model 311/611 Incubator that you have received meets its published specification.
Section 3 SPECIFICATIONS Temperature Range Ambient +5 to 95C Temperature Uniformity 311D +0.2 @ 37C 611D +0.25 @ 37C Temperature Accura
Section 4 RECEIVING AND INSPECTION 4.1 Inspection: The carrier, when accepting shipment, also accepts responsibility for safe delivery and is liabl
Section 5 INSTALLATION Local city, county, or other ordinances may govern the use of this equipment. If you have any questions about local requireme
85.5 Cleaning: The unit chamber should be cleaned and disinfected before use. Remove all of the interior parts, and clean thoroughly with a
Section 6 CONTROLS OVERVIEW 6.1 Power Switch: The main power ON/OFF switch controls all power to the unit. It must be on before any systems are oper