HIGH PERFORMANCE INCUBATORS MODELS: 311D, 611D INSTALLATION AND INSTRUCTION MANUAL

Seite 2 - TABLE OF CONTENTS

Section 7 OPERATION 7.1 Check power supply against unit serial plate. They must match. 7.2 Plug service cord into the electrical outlet. Push pow

Seite 3 - INTRODUCTION

11stable temperature the fan may be turned off.

Seite 4 - WARRANTY

CLEANING AND ROUTINE CARE Section 8 Both the interior and exterior of the incubator should be wiped down with a soft cloth periodically to prevent a

Seite 5 - SPECIFICATIONS

SERVICE AND TROUBLE-SHOOTING Section 9 TEMPERATURE Temp-1 Temperature too high-display and reference thermometer don’t match 1/ Controller

Seite 6 - RECEIVING AND INSPECTION

14 1/ Safety set too low – see section 7.5. 2/ Controller set too low – see section 7.3. 3/ Unit not recovered from door opening – wait for di

Seite 7 - INSTALLATION

15including motors, arcing relays or radio transmitters. 7/ Bad connection on temperature sensor or faulty sensor – check connectors for continuit

Seite 8

16 1/ If shaft spins freely: check connections to motor and check voltage to motor. 2/ If shaft rubs or is frozen, relieve binding and retest.

Seite 9 - CONTROLS OVERVIEW

17 OTHER Other-1 Controller on at all times - "locked-up" 1/ Turn unit off and on to reset. 2/ If

Seite 10 - OPERATION

18 PARTS LIST Description 120V 220V Blower Motor I-104100 I-210001 Circuit Breaker I-1100505 I-1100505 Convenience Outlet I-100020 I

Seite 11

2TABLE OF CONTENTS SECTION 1.0 INTRODUCTION SECTION 2.0 WARRANTY SECTION 3.0 SPECIFICATIONS SECTION 4.0 RECEIVING

Seite 12 - CLEANING AND ROUTINE CARE

Section 1 INTRODUCTION The Labnet Incubators have been designed for general purpose incubations (<65ºC), as well as those applications requiring h

Seite 13 - TEMPERATURE

Section 2 WARRANTY Labnet International, Inc. guarantees that the Model 311/611 Incubator that you have received meets its published specification.

Seite 14

Section 3 SPECIFICATIONS Temperature Range Ambient +5 to 95C Temperature Uniformity 311D +0.2 @ 37C 611D +0.25 @ 37C Temperature Accura

Seite 15 - MECHANICAL

Section 4 RECEIVING AND INSPECTION 4.1 Inspection: The carrier, when accepting shipment, also accepts responsibility for safe delivery and is liabl

Seite 16

Section 5 INSTALLATION Local city, county, or other ordinances may govern the use of this equipment. If you have any questions about local requireme

Seite 17

85.5 Cleaning: The unit chamber should be cleaned and disinfected before use. Remove all of the interior parts, and clean thoroughly with a

Seite 18 - PARTS LIST

Section 6 CONTROLS OVERVIEW 6.1 Power Switch: The main power ON/OFF switch controls all power to the unit. It must be on before any systems are oper