MultiGene Mini Thermal Cycler TC020-24 /TC020-24-230V TC050-18 /TC050-18-230V Operation Manual Ver.1.2 Labnet International
8 user. Shift to the File Name Column, using the “right” arrow button, then select the user’s file using the up/down keys. Press “Edit” to ed
9 Hotlid=020C 2008-11-01 10:05:00 100 75 50 25 0 Temp 1c 95.0 2c 55.0 3c 72.0 4 04.0 Time 01:00
10 File Saving Screen In this screen use the keypad to enter the file name. Press “Enter” to move the cursor down one position. Sele
11 Enter the sample volume. Press “Back” to go back to the file save menu. Press “Run” to start running the file. In the file save menu, if
12 1.2 How to Set System Parameters In the main screen press “System” to enter the system parameter settings menu. Press “Enter” to move
13 “Tube mode” controls the cooling & heating rate of the block, inconsideration of the sample temperature changing at a slower rate. In tube mo
14 CHAPTER 4 TROUBLE SHOOTING 1 Failure Analysis and Processing Procedures No. Phenomenon Failure Analysis Troubleshooting Disconnected power s
15 2 Error Messages No. Error Message Possible Reasons 1 Error Message when running “Please contact the distributor”, with “AD1213 failure” The mi
16 APPENDIX 1 Wiring Diagram (1 of 2) LCD11LCDHeatersinkAC input_220-240VAC50/60HzACNSwitchF USE25* 20m m _ 2.5ASW1PEF USE15* 20m m _ 2.5ABLOCKP
17 Wiring Diagram (2 of 2) NPOWERPCB3BYQ504104200000UEJ2-1_HV+EJ2-2_HV-EJ2-3_+5VEJ2-4_HOTLIDEJ2-5_F ANEJ2-6_PWMEJ2-7_HEATEJ2-8_COOLEJ4-1_F AN-EJ4-
CONTENTS CHAPTER 1 SPECIFICATIONS ... 1 1 Norm
18 APPENDIX 2 Parameter Values Item Parameter Scale Range Note 1 Temp(ºC) Temp. Control point 4.0~99.0 --:-- (set the value before “:”through “Fu
19 Image 2 Image1 block reagent block reagent
20 Limited Warranty Labnet International, Inc. warrants that this product will be free from defects in material and workmanship for a period of 18
1 CHAPTER 1 SPECIFICATIONS This chapter describes the MultiGene Mini Thermal Cycler’s operation, transportation and storage conditions, basic pa
2 4 Specifications Temperature range: 4ºC~99ºC Maximum Heating rate: ≥5.0ºC/sec(max) Maximum Cooling rate:
3 CHAPTER 2 Features This chapter covers the MultiGene Mini Thermal Cycler’s mechanics, the keyboard and each key’s function, and preparations befo
4 2 Keyboard Diagram 3 Key Functions “CE”,can be used to change “+, -” when editing temperature a
5 4 Inspection before Power-up Before switching the power on, please ensure that: 1) Supply voltage matches the voltage rating label on the instr
6 CHAPTER 3 OPERATION GUIDE There are eight menu selections for the MultiGene Mini Thermal Cycler. In this chapter, we will explain op
7 Hotlid=020C 2009-03-24 10:05:00 Labnet MultiGene Mini Version: 2009-1.0 Default User: Labnet Default File: